NPPTL Respirator Assessments to Support the COVID-19 Response

Updated April 16, 2021

Important Update

As a consequence of the removal of these crisis capacity strategies, NPPTL will no longer conduct any of the Respirator Assessments to Support the COVID-19 Response.

Decontaminated Assessment Results

NPPTL has completed Decontaminated Assessments for the products listed below.

While decontamination and reuse of FFRs are not consistent with standard and approved usage, these options may need to be considered when FFR shortages exist. This assessment was developed to quantify the filtration efficiency and manikin fit factor of an N95 respirator that has been decontaminated. This assessment is not to determine the effectiveness of the decontamination procedure at killing pathogenic microorganisms. Respirators with exhalation valves are included in this assessment and are evaluated as received by NPPTL for manikin fit testing. Exhalation valves are sealed with wax for filtration testing and therefore, the performance of the exhalation valve is not examined for the filtration test. The results provided in this report are specific to the subset of samples that were provided to NPPTL for assessment. These results may be used to update the CDC guidance for Crisis Capacity Strategies (during known shortages).

Decontaminated Assessment Results
Respirator Model Decon.
Method
Decon.
Cycles
Filtration Efficiency Results Strap
Performance
Manikin
Fit Factor
Test
Report
Max. % Min. %
3M 1860 ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs 4 99.73 98.82 No visual degradation
> 15% change in force
Discolored nosepiece foam
≥100 Report 3
3M 1860 DiKlor-G® Sterilization 6 99.62 99.13 No visual degradation. >10% change in force
Discolored nosepiece foam
≥ 100 Report 10
3M 8000* Gas phase chlorine dioxide at
1000~2000 ppm·hr
5 Not assessed Not assessed Discolored straps
<10% change in force
≥ 100 Report 11
3M 8000* Gas phase chlorine dioxide at
1000~2000 ppm·hr
10 Not assessed Not assessed Discolored straps
<15% change in force
≥ 100 Report 11
3M 8000 Gas phase chlorine dioxide at
1000~2000 ppm·hr
20 99.30 98.84 Discolored straps
<10% change in force
Min: 92
Max: ≥100
Report 11
3M 8200* Chlorine Dioxide at 2,500 ppm·hours 4 99.59 99.39 >10% change in force
Discolored nosepiece foam and straps
Not assessed Report 31
3M 8200 Chlorine dioxide gas,  5 x 850 ppm-hours 5 99.52 98.74 >10% change in force
Discolored nosepiece foam and straps
≥100 Report 36
3M 8210 DiKlor-G® Sterilization 6 99.78 98.73 No visual degradation
>15% change in force
Discolored nosepiece foam
≥ 100 Report 4
3M V-flex 1805 ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs 1 99.80 99.50 No visual degradation
<10% change in force
≥100 Report 3
VWR 89201 ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs 4 99.59 97.91 No visual degradation
<10% change in force
Min: 37
Max: 96
Report 3
3M VFlex 1804* Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min 1 91.79 82.00 Respirators were soiled and information blurred/illegible
>10% change in force
≥100 Report 39
3M 1860 Supercritical CO2 10 89.90 66.90 Discolored respirator surface
<5% change in force
Min: 54
Max: ≥100
Report 20
3M 1860* Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min 1 74.70 55.70 Respirators were soiled and information blurred/illegible
>5% change in force
Min: 22
Max: 42
Report 39
3M 1860S* Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min 1 84.90 57.70 Respirators were soiled and information blurred/illegible
>10% change in force
Min: 8
Max: 13
Report 39
3M Vflex 1804 Dry Heat, 75-78°C recirculating hot air for 45 minutes 10 99.85 98.98 Respirator information on front of respirator was blurred/illegible
<5% change in force
Min: 70
Max: ≥100
Report 37
3M VFlex 1804 Dry Heat, 75-78 °C, 45 mins/cycle 10 99.91 99.72 Respirator information on front of respirator was blurred/illegible
>10% change in force
≥100 Report 44
3M 1860 Commercial laundry tumbler dryers,
70°C for 60 min
3 99.33 98.53 No visual degradation
<5% change in force
≥100 Report 12
3M 1860 Commercial laundry tumbler dryers,
>80°C for >70 min
3 98.89 95.08 No visual degradation
>5% change in force
≥100 Report 18
3M 1860* Dry Heat , 75°C for 60 min per cycle 5 99.02 98.32 No visual degradation
<5% change in force
≥100 Report 21
3M 1860 Dry Heat, 75-78°C recirculating hot air for 45 minutes 10 98.96 97.70 Respirator information on front of respirator was blurred/illegible
>5% change in force
≥100 Report 37
3M 1860 MSU Extension Dry Heat Decon (85 °C for 50 minutes) 4 99.02 98.14 Discolored straps.  Respirator information on front of respirator was blurred/illegible
>15% change in force
Min: 64
Max: ≥100
Report 40
3M 1860S Dry Heat, 75-78°C recirculating hot air for 45 minutes 10 98.79 97.86 Respirator information on front of respirator was blurred/illegible
>5% change in force
≥100 Report 37
3M 1870+* Dry Heat , 75°C for 60 min per cycle 5 99.81 98.60 No visual degradation
<5% change in force
≥100 Report 21
3M Aura 1870+ Dry Heat, 75-78°C recirculating hot air for 45 minutes 10 99.98 98.95 <5% change in force ≥100 Report 37
3M 8210 MSU Extension Dry Heat Decon (85 °C for 50 minutes) 4 99.96 97.33 Respirator information on front of respirator was blurred/illegible
>40% change in force
≥100 Report 40
3M 8210+ Dry heat, 75°C for 30 min 10 99.45 98.66 No visual degradation
<5% change in force
≥100 Report 25
3M 8511 Commercial laundry tumbler dryers,
70°C for 60 min
3 99.82 99.09 No visual degradation
<5% change in force
≥100 Report 12
3M 8511 Commercial laundry tumbler dryers,
>80°C for >70 min
3 99.12 98.14 No visual degradation
>10% change in force
≥100 Report 18
3M Aura 9205+ Dry Heat, 75-78°C recirculating hot air for 45 minutes 10 99.91 99.39 >5% change in force ≥100 Report 37
Halyard 46727* Dry Heat , 75°C for 60 min per cycle 5 99.54 98.76 No visual degradation
<5% change in force
≥100 Report 21
Halyard 62126 Commercial laundry tumbler dryers,
70°C for 60 min
3 99.78 99.18 No visual degradation
<5% change in force
≥100 Report 12
Halyard 62126 Commercial laundry tumbler dryers,
>80°C for >70 min
3 99.32 97.31 No visual degradation
>5% change in force
≥100 Report 18
Halyard Fluidshield 46767 Regular The Clean Sleep Hyperion, 90 mins at 175°F 5 93.77 61.90 <5% change in force
Respirators were significantly damaged.
Min: 2
Max: 5
Report 32
Halyard Fluidshield 46767 Regular Clean Sleep Hyperion, 90 mins at 175°F 3 99.62 97.89 No visual degradation
<5% change in force
≥100 Report 35
3M 1870* Gaseous ozone, 450 ppm 5 99.93 99.72 No visual degradation
>5% change in force
≥100 Report 6
3M 1860* Gravity Steam Cycle 3 98.24 96.93 No visual degradation
>5% change in force
Min: 18
Max: 20
Report 15
3M 1860* Cres Cor moist heat decontamination system:
1 hour of exposure at 100°C and 100% RH
1 98.02 97.38 No visual degradation.
>5% change in force.
Respirator information on front of respirator was blurred/illegible.  
Min: 20
Max: 92
Report 26
3M 1860* Cres Cor moist heat decontamination system:
1 hour of exposure at 100°C and 100% RH
3 98.45 97.24 No visual degradation of the straps. >10% change in force.
Respirator information on front of respirator was blurred/illegible.  
Min: 15
Max: 38
Report 26
3M 1860* Cres Cor moist heat decontamination system:
1 hour of exposure at 100°C and 100% RH
5 98.08 96.92 No visual degradation of the straps. >10% change in force.
Respirator information on front of respirator was blurred/illegible.  
Min: 24
Max: 51
Report 26
3M 1860* Cres Cor moist heat decontamination cabinet,
15 mins at 85°C and 65% RH
1 99.34 99.08 <5% change in force
Blurred print on front of respirators
Min: 82
Max: 114
Report 33
3M 1860* Cres Cor moist heat decontamination cabinet,
15 mins at 85°C and 65% RH
3 99.25 99.05 <5% change in force
Blurred print on front of respirators
Min: 86
Max: 108
Report 33
3M 1860* Cres Cor moist heat decontamination cabinet,
15 mins at 85°C and 65% RH
5 98.92 98.04 <5% change in force
Blurred print on front of respirators
Min: 93
Max: 116
Report 33
3M 1870* Gravity Steam Cycle 3 98.59 97.49 No visual degradation
>20% change in force
≥100 Report 15
3M 8210 Electric Steamer, 100°C for 30 sec/cycle 10 99.46 98.03 Respirator information on front of respirator was blurred/illegible
>10% change in force
Min: 18
Max: 22
Report 38
3M V-flex 9105* Gravity Steam Cycle 3 98.99 98.47 No visual degradation
>10% change in force
≥100 Report 15
3M 8200* High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. 1 55.7 46.6 No visual degradation
<5% change in force
Not assessed Report 5
3M 8200* High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. 2 51.1 47.2 No visual degradation
>10 % change in force
Not assessed Report 5
3M 8200* High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. 3 50.6 47.4 No visual degradation
<15% change in force
Not assessed Report 5
3M 8200* Electron Beam Irradiation 1 63.6 No visual degradation.
>10% change in force
Not assessed Report 27
N95 3M 1860S Methylene Blue 5 98.65 93.59 No visual degradation
<5% change in force
≥100 Report 1
3M 1860*

Methylene Blue + Light

5 99.00 98.34 No visual degradation
<5% change in force
≥100 Report 21
3M 1870+*

Methylene Blue + Light

5 99.88 99.64 No visual degradation
<10% change in force
≥100 Report 21
Halyard 46727*

Methylene Blue + Light

5 99.44 97.24 No visual degradation
<5% change in force
≥100 Report 21
3M 1860* Microwave generated plasma 3 99.50 98.81 No visual degradation
>5% change in force
≥100 Report 24
3M 8210* Microwave generated plasma 3 99.33 91.54 No visual degradation
>5% change in force
Min: 83
Max: ≥100
Report 19
3M 8511 Sterrad NX100 Hydrogen Peroxide
Vapor/Low Temp plasma
2 78.90 58.30 No visual degradation
<5% change in force
Min: 40
Max: ≥100
Report 16
Halyard Fluidshield 46727 Reg Sterrad NX100 Hydrogen Peroxide
Vapor/Low Temp plasma
2 98.82 98.29 No visual degradation
>5% change in force
≥100 Report 8
3M 1860* UV-Concepts Inc. model UVE (UV-C Enclosure). 180 seconds and produces ~1000 mJ/cm² of UV-C “dose.” 10 99.61 98.97 No visual degradation
<5% change in force.
>100 Report 30
3M 1860S* Broadband UVC (DECON95) 1 99.24 98.58 No visual degradation.
<5% change in force
Not assessed Report 29
3M 1860S* Broadband UVC (DECON95) 3 99.22 97.96 No visual degradation.
<5% change in force
Not assessed Report 29
3M 1860S* Broadband UVC (DECON95) 5 98.89 98.12 No visual degradation.
>5% change in force
Not assessed Report 29
3M 8210* UV-C 254 nm –
500 mJ/cm
2
10 99.61 99.15 No visual degradation
>5% change in force
≥100 Report 22
3M 8210* UV-C 254 nm –
1000 mJ/cm2
10  99.41  99.10 No visual degradation
>5% change in force
≥100 Report 22
3M 8210 UV-C dose ˃ 2 J/cm² for 1 min/cycle 20 99.70 98.37 No visual degradation
>10% change in force
≥100 Report 42
Moldex 1513 Large UV-C dose ˃ 2 J/cm² for 1 min/cycle 20 99.97 98.38 Separated nosepiece foam on 1 sample  <5% change in force Min: 16
Max: ≥100
Report 41
3M 1860* 1.06% peracetic acid (PAA) solution 1 99.02 98.58 No visual degradation.
<5% change in force
>100 Report 28
3M 1860* 1.06% peracetic acid (PAA) solution 3 99.49 99.08 No visual degradation.
<5% change in force
>100 Report 28
3M 1860* 1.06% peracetic acid (PAA) solution 5 99.29 98.64 No visual degradation.
<5% change in force
>100 Report 28
BYD DE2322* Plasma Discharge ROS 1 98.04 97.79 No visual degradation.
<5% change in force
Not assessed Report 27
BYD DE2322* Plasma Discharge ROS 3 98.69 97.10 No visual degradation.
<5% change in force
Not assessed Report 27
Prestige Ameritech RP88020* Plasma Discharge ROS 1 99.56 99.35 No visual degradation.
<5% change in force
Not assessed Report 27
Prestige Ameritech RP88020* Plasma Discharge ROS 3 99.59 99.50 No visual degradation.
<5% change in force
Not assessed Report 27
3M 1860* VPHP 5 99.54 99.26 No control provided ≥100 Report 2
3M 1860 VPHP 10 99.47 99.02 No visual degradation
≤ 35% change in force
≥100 Report 7
3M 1860 VPHP 15 99.75 99.32 No visual degradation
>40% change in force
≥100 Report 14
3M 1860 VPHP 20 99.74 99.16 No visual degradation
>35% change in force
≥100 Report 14
3M 1860 Bioquell HPV, >480 ppm HPV at 900 sec/cycle 5 99.37 98.67 No visual degradation
>15% change in force
≥100 Report 43
3M 1860* VHP 5 98.97 97.67 Discolored nosepiece foam
>30% change in force
≥100 Report 21
3M 1860S* VPHP 5 99.73 99.01 No visual degradation
< 15% change in force
≥100 Report 2
3M 1870+* VPHP 20 99.78 98.00 No visual degradation
>15% change in force
≥100 Report 17
3M 1870+* VHP 5 99.89 99.52 No visual degradation
>10% change in force
≥100 Report 21
3M 8000* VHP 7% 1 98.99 98.69 No control provided Not assessed Report 9
3M 8000* VHP 7% 2 99.05 98.33 No control provided Not assessed Report 9
3M 8000* VHP 7% 3 98.80 98.00 No control provided Not assessed Report 9
3M 8000* VHP 7% 4 99.18 98.74 No control provided Not assessed Report 9
3M 8000* VHP 7% 5 99.16 98.78 No control provided Not assessed Report 9
3M 8210* VPHP 5 99.72 99.48 No visual degradation
> 10% change in force
≥100 Report 2
3M 8210 VPHP 10 99.60 99.37 No visual degradation
≤ 10% change in force
≥100 Report 7
3M 8210* VHP 7% 1 99.53 97.26 No control provided ≥100 Report 9
3M 8210* VHP 7% 2 99.41 99.36 No control provided ≥100 Report 9
3M 8210* VHP 7% 3 99.50 97.60 No control provided ≥100 Report 9
3M 8210* VHP 7% 4 99.52 99.31 No control provided ≥100 Report 9
3M 8210* VHP 7% 5 99.49 99.09 No control provided ≥100 Report 9
3M 8210 VPHP 20 99.90 98.19 No visual degradation
>15% change in force
≥100 Report 14
3M 8210V* VHP 7% 1 98.94 98.24 No control provided ≥100 Report 9
3M 8210V* VHP 7% 2 98.79 98.57 No control provided ≥100 Report 9
3M 8210V* VHP 7% 3 98.53 98.30 No control provided ≥100 Report 9
3M 8210V* VHP 7% 4 99.10 97.97 No control provided ≥100 Report 9
3M 8210V* VHP 7% 5 98.62 98.34 No control provided ≥100 Report 9
3M 8511 VPHP 5 99.81 99.63 No visual degradation
>20% change in force
≥100 Report 14
3M 8511 VPHP 10 99.88 99.46 No visual degradation
>20% change in force
≥100 Report 14
3M 8511 VPHP 20 99.73 99.26 No visual degradation
>15% change in force
≥100 Report 13
3M 9205+ VPHP 5 99.86 98.11 No visual degradation
<15% change in force
≥100 Report 23
3M 9205+ VPHP 10 99.78 97.77 No visual degradation
<15% change in force
≥100 Report 23
3M 9205+ VPHP 20 99.69 96.13 No visual degradation
<15% change in force
≥100 Report 23
3M V-flex 1804* VPHP 5 99.71 99.50 No visual degradation
>5% change in force
≥100 Report 2
3M V-flex 1804 VPHP 10 99.59 99.28 No visual degradation
≤ 25% change in force
Min: 27
Max: ≥100
Report 7
3M V-Flex 1804 VPHP 20 99.67 99.00 Discolored straps
>5% change in force
Min: 85
Max: ≥100
Report 14
Crosstex GPRN95* VHP 7% 1 99.66 99.09 No control provided Not assessed Report 9
Crosstex GPRN95* VHP 7% 2 99.62 99.03 No control provided Not assessed Report 9
Crosstex GPRN95* VHP 7% 3 99.28 97.85 No control provided Not assessed Report 9
Crosstex GPRN95* VHP 7% 4 99.22 96.78 No control provided Not assessed Report 9
Crosstex GPRN95* VHP 7% 5 99.13 96.84 No control provided Not assessed Report 9
Gerson 1730 VPHP 5 99.65 99.26 No visual degradation.
<5% change in force
≥100 Report 17
Gerson 1730 VPHP 10 99.44 97.77 No visual degradation
>15% change in force
≥100 Report 17
Gerson 1730 VPHP 20 99.48 99.20 No visual degradation
>5% change in force
≥100 Report 17
Gerson 1740 VPHP 5 99.35 98.82 No visual degradation
>10% change in force
Min: 88
Max: ≥100
Report 17
Gerson 1740 VPHP 10 99.40 98.47 No visual degradation
>5% change in force
Min: 68
Max: ≥100
Report 17
Gerson 1740 VPHP 20 99.36 98.58 No visual degradation
>5% change in force
≥100 Report 17
Halyard 46727* VHP 5 99.33 97.89 No visual degradation
>15% change in force
≥100 Report 21
Halyard 46727 Bioquell HPV, >480 ppm HPV at 900 sec/cycle 5 99.74 98.44 No visual degradation
<5% change in force
≥100 Report 43
Makrite 9500-N95S* H₂O₂ (35% concentration) + catalyst 10 99.46 96.15 >5% change in force
Discolored nosepiece foam.
≥100 Report 34
Makrite 9500-N95S* H₂O₂ (35% concentration) + catalyst 20 99.55 85.60 <5% change in force
Discolored nosepiece foam
Min: 44
Max: ≥100
Report 34
Makrite 9500-N95S* H₂O₂ (35% concentration) + catalyst 30 99.54 98.10 >10% change in force
Discolored nosepiece foam
Min: 28
Max: 88
Report 34
Moldex 1512* VHP 7% 1 98.79 97.91 No control provided ≥100 Report 9
Moldex 1512* VHP 7% 2 99.05 98.74 No control provided ≥100 Report 9
Moldex 1512* VHP 7% 3 98.59 97.80 No control provided ≥100 Report 9
Moldex 1512* VHP 7% 4 98.78 97.76 No control provided ≥100 Report 9
Moldex 1512* VHP 7% 5 98.74 97.55 No control provided ≥100 Report 9
Moldex 1512 VPHP 20 99.53 97.56 No visual degradation
>15% change in force
≥100 Report 14
Moldex 1512 Medium* VPHP 5 99.12 96.25 No visual degradation
> 10% change in force
≥100 Report 2
Moldex 1512 Medium VPHP 10 99.40 95.19 No visual degradation
≤ 10% change in force
≥100 Report 7
Moldex 2200 VPHP 20 99.46 99.38 No visual degradation
<5% change in force
Min: 90
Max: ≥100
Report 14
Moldex 2200 M/L* VPHP 5 99.25 97.54 No visual degradation
≤ 5% change in force
Min: 34
Max: ≥100
Report 2
Moldex 2200 M/L VPHP 10 98.76 94.27 No visual degradation
≤ 10% change in force
≥100 Report 7
Sperian N1105 VPHP 10 99.88 99.74 No visual degradation
>10% change in force
≥100 Report 13
Sperian N1105 VPHP 20 99.94 99.75 No visual degradation
>10% change in force
Min: 58
Max: ≥100
Report 14
Sperian N1105 M/L VPHP 5 99.85 99.54 No visual degradation
≤ 20% change in force
≥100 Report 7
Sperian N1125 VPHP 10 99.83 99.24 No visual degradation
>5% change in force
Min: 89
Max: ≥100
Report 14
Sperian N1125 VPHP 20 99.85 99.33 No visual degradation
>5% change in force
Min: 58
Max: ≥100
Report 14
Sperian N1125 M/L VPHP 5 99.88 99.13 No visual degradation
≤ 15% change in force
Min: 99
Max: ≥100
Report 7
Sperian One-Fit VPHP 5 99.75 98.90 No visual degradation
≤ 5% change in force
Not assessed Report 7
Sperian One-fit VPHP 10 99.80 97.01 No visual degradation
>15% change in force
Not assessed Report 14
Sperian One-fit VPHP 20 99.85 99.42 No visual degradation
>15% change in force
Not assessed Report 14

* Testing for this model did not meet the requirements of the Decontaminated Respirator Assessment Plan because an insufficient number of samples were provided to NIOSH for 1 or more tests. Therefore, results are posted for the number of samples available. Please see results report for actual number of samples evaluated for filtration efficiency, manikin fit, and strap performance.