NPPTL Respirator Assessments to Support the COVID-19 Response
Updated April 16, 2021
As a consequence of the removal of these crisis capacity strategies, NPPTL will no longer conduct any of the Respirator Assessments to Support the COVID-19 Response.
Decontaminated Assessment Results
NPPTL has completed Decontaminated Assessments for the products listed below.
While decontamination and reuse of FFRs are not consistent with standard and approved usage, these options may need to be considered when FFR shortages exist. This assessment was developed to quantify the filtration efficiency and manikin fit factor of an N95 respirator that has been decontaminated. This assessment is not to determine the effectiveness of the decontamination procedure at killing pathogenic microorganisms. Respirators with exhalation valves are included in this assessment and are evaluated as received by NPPTL for manikin fit testing. Exhalation valves are sealed with wax for filtration testing and therefore, the performance of the exhalation valve is not examined for the filtration test. The results provided in this report are specific to the subset of samples that were provided to NPPTL for assessment. These results may be used to update the CDC guidance for Crisis Capacity Strategies (during known shortages).
Respirator Model | Decon. Method |
Decon. Cycles |
Filtration Efficiency Results | Strap Performance |
Manikin Fit Factor |
Test Report |
|
---|---|---|---|---|---|---|---|
Max. % | Min. % | ||||||
3M 1860 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 4 | 99.73 | 98.82 | No visual degradation > 15% change in force Discolored nosepiece foam |
≥100 | Report 3 |
3M 1860 | DiKlor-G® Sterilization | 6 | 99.62 | 99.13 | No visual degradation. >10% change in force Discolored nosepiece foam |
≥ 100 | Report 10 |
3M 8000* | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
5 | Not assessed | Not assessed | Discolored straps <10% change in force |
≥ 100 | Report 11 |
3M 8000* | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
10 | Not assessed | Not assessed | Discolored straps <15% change in force |
≥ 100 | Report 11 |
3M 8000 | Gas phase chlorine dioxide at 1000~2000 ppm·hr |
20 | 99.30 | 98.84 | Discolored straps <10% change in force |
Min: 92 Max: ≥100 |
Report 11 |
3M 8200* | Chlorine Dioxide at 2,500 ppm·hours | 4 | 99.59 | 99.39 | >10% change in force Discolored nosepiece foam and straps |
Not assessed | Report 31 |
3M 8200 | Chlorine dioxide gas, 5 x 850 ppm-hours | 5 | 99.52 | 98.74 | >10% change in force Discolored nosepiece foam and straps |
≥100 | Report 36 |
3M 8210 | DiKlor-G® Sterilization | 6 | 99.78 | 98.73 | No visual degradation >15% change in force Discolored nosepiece foam |
≥ 100 | Report 4 |
3M V-flex 1805 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 1 | 99.80 | 99.50 | No visual degradation <10% change in force |
≥100 | Report 3 |
VWR 89201 | ClorDiSys Solutions Minidox-M, Chlorine Dioxide Generator, Concentration: 1 mg/L, RH: 65%, Exposure: 720 ppm/hrs | 4 | 99.59 | 97.91 | No visual degradation <10% change in force |
Min: 37 Max: 96 |
Report 3 |
3M VFlex 1804* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 91.79 | 82.00 | Respirators were soiled and information blurred/illegible >10% change in force |
≥100 | Report 39 |
3M 1860 | Supercritical CO2 | 10 | 89.90 | 66.90 | Discolored respirator surface <5% change in force |
Min: 54 Max: ≥100 |
Report 20 |
3M 1860* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 74.70 | 55.70 | Respirators were soiled and information blurred/illegible >5% change in force |
Min: 22 Max: 42 |
Report 39 |
3M 1860S* | Dense-phase CO2, 37°C above its supercritical pressure of 1100 psig with humidification for 90 min | 1 | 84.90 | 57.70 | Respirators were soiled and information blurred/illegible >10% change in force |
Min: 8 Max: 13 |
Report 39 |
3M Vflex 1804 | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.85 | 98.98 | Respirator information on front of respirator was blurred/illegible <5% change in force |
Min: 70 Max: ≥100 |
Report 37 |
3M VFlex 1804 | Dry Heat, 75-78 °C, 45 mins/cycle | 10 | 99.91 | 99.72 | Respirator information on front of respirator was blurred/illegible >10% change in force |
≥100 | Report 44 |
3M 1860 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.33 | 98.53 | No visual degradation <5% change in force |
≥100 | Report 12 |
3M 1860 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 98.89 | 95.08 | No visual degradation >5% change in force |
≥100 | Report 18 |
3M 1860* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.02 | 98.32 | No visual degradation <5% change in force |
≥100 | Report 21 |
3M 1860 | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 98.96 | 97.70 | Respirator information on front of respirator was blurred/illegible >5% change in force |
≥100 | Report 37 |
3M 1860 | MSU Extension Dry Heat Decon (85 °C for 50 minutes) | 4 | 99.02 | 98.14 | Discolored straps. Respirator information on front of respirator was blurred/illegible >15% change in force |
Min: 64 Max: ≥100 |
Report 40 |
3M 1860S | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 98.79 | 97.86 | Respirator information on front of respirator was blurred/illegible >5% change in force |
≥100 | Report 37 |
3M 1870+* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.81 | 98.60 | No visual degradation <5% change in force |
≥100 | Report 21 |
3M Aura 1870+ | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.98 | 98.95 | <5% change in force | ≥100 | Report 37 |
3M 8210 | MSU Extension Dry Heat Decon (85 °C for 50 minutes) | 4 | 99.96 | 97.33 | Respirator information on front of respirator was blurred/illegible >40% change in force |
≥100 | Report 40 |
3M 8210+ | Dry heat, 75°C for 30 min | 10 | 99.45 | 98.66 | No visual degradation <5% change in force |
≥100 | Report 25 |
3M 8511 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.82 | 99.09 | No visual degradation <5% change in force |
≥100 | Report 12 |
3M 8511 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 99.12 | 98.14 | No visual degradation >10% change in force |
≥100 | Report 18 |
3M Aura 9205+ | Dry Heat, 75-78°C recirculating hot air for 45 minutes | 10 | 99.91 | 99.39 | >5% change in force | ≥100 | Report 37 |
Halyard 46727* | Dry Heat , 75°C for 60 min per cycle | 5 | 99.54 | 98.76 | No visual degradation <5% change in force |
≥100 | Report 21 |
Halyard 62126 | Commercial laundry tumbler dryers, 70°C for 60 min |
3 | 99.78 | 99.18 | No visual degradation <5% change in force |
≥100 | Report 12 |
Halyard 62126 | Commercial laundry tumbler dryers, >80°C for >70 min |
3 | 99.32 | 97.31 | No visual degradation >5% change in force |
≥100 | Report 18 |
Halyard Fluidshield 46767 Regular | The Clean Sleep Hyperion, 90 mins at 175°F | 5 | 93.77 | 61.90 | <5% change in force Respirators were significantly damaged. |
Min: 2 Max: 5 |
Report 32 |
Halyard Fluidshield 46767 Regular | Clean Sleep Hyperion, 90 mins at 175°F | 3 | 99.62 | 97.89 | No visual degradation <5% change in force |
≥100 | Report 35 |
3M 1870* | Gaseous ozone, 450 ppm | 5 | 99.93 | 99.72 | No visual degradation >5% change in force |
≥100 | Report 6 |
3M 1860* | Gravity Steam Cycle | 3 | 98.24 | 96.93 | No visual degradation >5% change in force |
Min: 18 Max: 20 |
Report 15 |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
1 | 98.02 | 97.38 | No visual degradation. >5% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 20 Max: 92 |
Report 26 |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
3 | 98.45 | 97.24 | No visual degradation of the straps. >10% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 15 Max: 38 |
Report 26 |
3M 1860* | Cres Cor moist heat decontamination system: 1 hour of exposure at 100°C and 100% RH |
5 | 98.08 | 96.92 | No visual degradation of the straps. >10% change in force. Respirator information on front of respirator was blurred/illegible. |
Min: 24 Max: 51 |
Report 26 |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
1 | 99.34 | 99.08 | <5% change in force Blurred print on front of respirators |
Min: 82 Max: 114 |
Report 33 |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
3 | 99.25 | 99.05 | <5% change in force Blurred print on front of respirators |
Min: 86 Max: 108 |
Report 33 |
3M 1860* | Cres Cor moist heat decontamination cabinet, 15 mins at 85°C and 65% RH |
5 | 98.92 | 98.04 | <5% change in force Blurred print on front of respirators |
Min: 93 Max: 116 |
Report 33 |
3M 1870* | Gravity Steam Cycle | 3 | 98.59 | 97.49 | No visual degradation >20% change in force |
≥100 | Report 15 |
3M 8210 | Electric Steamer, 100°C for 30 sec/cycle | 10 | 99.46 | 98.03 | Respirator information on front of respirator was blurred/illegible >10% change in force |
Min: 18 Max: 22 |
Report 38 |
3M V-flex 9105* | Gravity Steam Cycle | 3 | 98.99 | 98.47 | No visual degradation >10% change in force |
≥100 | Report 15 |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 1 | 55.7 | 46.6 | No visual degradation <5% change in force |
Not assessed | Report 5 |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 2 | 51.1 | 47.2 | No visual degradation >10 % change in force |
Not assessed | Report 5 |
3M 8200* | High energy (10 MeV), high dose rate (3 kGy/s) electron beam irradiation at 25 kGy per decontamination cycle. | 3 | 50.6 | 47.4 | No visual degradation <15% change in force |
Not assessed | Report 5 |
3M 8200* | Electron Beam Irradiation | 1 | 63.6 | – | No visual degradation. >10% change in force |
Not assessed | Report 27 |
N95 3M 1860S | Methylene Blue | 5 | 98.65 | 93.59 | No visual degradation <5% change in force |
≥100 | Report 1 |
3M 1860* |
Methylene Blue + Light |
5 | 99.00 | 98.34 | No visual degradation <5% change in force |
≥100 | Report 21 |
3M 1870+* |
Methylene Blue + Light |
5 | 99.88 | 99.64 | No visual degradation <10% change in force |
≥100 | Report 21 |
Halyard 46727* |
Methylene Blue + Light |
5 | 99.44 | 97.24 | No visual degradation <5% change in force |
≥100 | Report 21 |
3M 1860* | Microwave generated plasma | 3 | 99.50 | 98.81 | No visual degradation >5% change in force |
≥100 | Report 24 |
3M 8210* | Microwave generated plasma | 3 | 99.33 | 91.54 | No visual degradation >5% change in force |
Min: 83 Max: ≥100 |
Report 19 |
3M 8511 | Sterrad NX100 Hydrogen Peroxide Vapor/Low Temp plasma |
2 | 78.90 | 58.30 | No visual degradation <5% change in force |
Min: 40 Max: ≥100 |
Report 16 |
Halyard Fluidshield 46727 Reg | Sterrad NX100 Hydrogen Peroxide Vapor/Low Temp plasma |
2 | 98.82 | 98.29 | No visual degradation >5% change in force |
≥100 | Report 8 |
3M 1860* | UV-Concepts Inc. model UVE (UV-C Enclosure). 180 seconds and produces ~1000 mJ/cm² of UV-C “dose.” | 10 | 99.61 | 98.97 | No visual degradation <5% change in force. |
>100 | Report 30 |
3M 1860S* | Broadband UVC (DECON95) | 1 | 99.24 | 98.58 | No visual degradation. <5% change in force |
Not assessed | Report 29 |
3M 1860S* | Broadband UVC (DECON95) | 3 | 99.22 | 97.96 | No visual degradation. <5% change in force |
Not assessed | Report 29 |
3M 1860S* | Broadband UVC (DECON95) | 5 | 98.89 | 98.12 | No visual degradation. >5% change in force |
Not assessed | Report 29 |
3M 8210* | UV-C 254 nm – 500 mJ/cm2 |
10 | 99.61 | 99.15 | No visual degradation >5% change in force |
≥100 | Report 22 |
3M 8210* | UV-C 254 nm – 1000 mJ/cm2 |
10 | 99.41 | 99.10 | No visual degradation >5% change in force |
≥100 | Report 22 |
3M 8210 | UV-C dose ˃ 2 J/cm² for 1 min/cycle | 20 | 99.70 | 98.37 | No visual degradation >10% change in force |
≥100 | Report 42 |
Moldex 1513 Large | UV-C dose ˃ 2 J/cm² for 1 min/cycle | 20 | 99.97 | 98.38 | Separated nosepiece foam on 1 sample <5% change in force | Min: 16 Max: ≥100 |
Report 41 |
3M 1860* | 1.06% peracetic acid (PAA) solution | 1 | 99.02 | 98.58 | No visual degradation. <5% change in force |
>100 | Report 28 |
3M 1860* | 1.06% peracetic acid (PAA) solution | 3 | 99.49 | 99.08 | No visual degradation. <5% change in force |
>100 | Report 28 |
3M 1860* | 1.06% peracetic acid (PAA) solution | 5 | 99.29 | 98.64 | No visual degradation. <5% change in force |
>100 | Report 28 |
BYD DE2322* | Plasma Discharge ROS | 1 | 98.04 | 97.79 | No visual degradation. <5% change in force |
Not assessed | Report 27 |
BYD DE2322* | Plasma Discharge ROS | 3 | 98.69 | 97.10 | No visual degradation. <5% change in force |
Not assessed | Report 27 |
Prestige Ameritech RP88020* | Plasma Discharge ROS | 1 | 99.56 | 99.35 | No visual degradation. <5% change in force |
Not assessed | Report 27 |
Prestige Ameritech RP88020* | Plasma Discharge ROS | 3 | 99.59 | 99.50 | No visual degradation. <5% change in force |
Not assessed | Report 27 |
3M 1860* | VPHP | 5 | 99.54 | 99.26 | No control provided | ≥100 | Report 2 |
3M 1860 | VPHP | 10 | 99.47 | 99.02 | No visual degradation ≤ 35% change in force |
≥100 | Report 7 |
3M 1860 | VPHP | 15 | 99.75 | 99.32 | No visual degradation >40% change in force |
≥100 | Report 14 |
3M 1860 | VPHP | 20 | 99.74 | 99.16 | No visual degradation >35% change in force |
≥100 | Report 14 |
3M 1860 | Bioquell HPV, >480 ppm HPV at 900 sec/cycle | 5 | 99.37 | 98.67 | No visual degradation >15% change in force |
≥100 | Report 43 |
3M 1860* | VHP | 5 | 98.97 | 97.67 | Discolored nosepiece foam >30% change in force |
≥100 | Report 21 |
3M 1860S* | VPHP | 5 | 99.73 | 99.01 | No visual degradation < 15% change in force |
≥100 | Report 2 |
3M 1870+* | VPHP | 20 | 99.78 | 98.00 | No visual degradation >15% change in force |
≥100 | Report 17 |
3M 1870+* | VHP | 5 | 99.89 | 99.52 | No visual degradation >10% change in force |
≥100 | Report 21 |
3M 8000* | VHP 7% | 1 | 98.99 | 98.69 | No control provided | Not assessed | Report 9 |
3M 8000* | VHP 7% | 2 | 99.05 | 98.33 | No control provided | Not assessed | Report 9 |
3M 8000* | VHP 7% | 3 | 98.80 | 98.00 | No control provided | Not assessed | Report 9 |
3M 8000* | VHP 7% | 4 | 99.18 | 98.74 | No control provided | Not assessed | Report 9 |
3M 8000* | VHP 7% | 5 | 99.16 | 98.78 | No control provided | Not assessed | Report 9 |
3M 8210* | VPHP | 5 | 99.72 | 99.48 | No visual degradation > 10% change in force |
≥100 | Report 2 |
3M 8210 | VPHP | 10 | 99.60 | 99.37 | No visual degradation ≤ 10% change in force |
≥100 | Report 7 |
3M 8210* | VHP 7% | 1 | 99.53 | 97.26 | No control provided | ≥100 | Report 9 |
3M 8210* | VHP 7% | 2 | 99.41 | 99.36 | No control provided | ≥100 | Report 9 |
3M 8210* | VHP 7% | 3 | 99.50 | 97.60 | No control provided | ≥100 | Report 9 |
3M 8210* | VHP 7% | 4 | 99.52 | 99.31 | No control provided | ≥100 | Report 9 |
3M 8210* | VHP 7% | 5 | 99.49 | 99.09 | No control provided | ≥100 | Report 9 |
3M 8210 | VPHP | 20 | 99.90 | 98.19 | No visual degradation >15% change in force |
≥100 | Report 14 |
3M 8210V* | VHP 7% | 1 | 98.94 | 98.24 | No control provided | ≥100 | Report 9 |
3M 8210V* | VHP 7% | 2 | 98.79 | 98.57 | No control provided | ≥100 | Report 9 |
3M 8210V* | VHP 7% | 3 | 98.53 | 98.30 | No control provided | ≥100 | Report 9 |
3M 8210V* | VHP 7% | 4 | 99.10 | 97.97 | No control provided | ≥100 | Report 9 |
3M 8210V* | VHP 7% | 5 | 98.62 | 98.34 | No control provided | ≥100 | Report 9 |
3M 8511 | VPHP | 5 | 99.81 | 99.63 | No visual degradation >20% change in force |
≥100 | Report 14 |
3M 8511 | VPHP | 10 | 99.88 | 99.46 | No visual degradation >20% change in force |
≥100 | Report 14 |
3M 8511 | VPHP | 20 | 99.73 | 99.26 | No visual degradation >15% change in force |
≥100 | Report 13 |
3M 9205+ | VPHP | 5 | 99.86 | 98.11 | No visual degradation <15% change in force |
≥100 | Report 23 |
3M 9205+ | VPHP | 10 | 99.78 | 97.77 | No visual degradation <15% change in force |
≥100 | Report 23 |
3M 9205+ | VPHP | 20 | 99.69 | 96.13 | No visual degradation <15% change in force |
≥100 | Report 23 |
3M V-flex 1804* | VPHP | 5 | 99.71 | 99.50 | No visual degradation >5% change in force |
≥100 | Report 2 |
3M V-flex 1804 | VPHP | 10 | 99.59 | 99.28 | No visual degradation ≤ 25% change in force |
Min: 27 Max: ≥100 |
Report 7 |
3M V-Flex 1804 | VPHP | 20 | 99.67 | 99.00 | Discolored straps >5% change in force |
Min: 85 Max: ≥100 |
Report 14 |
Crosstex GPRN95* | VHP 7% | 1 | 99.66 | 99.09 | No control provided | Not assessed | Report 9 |
Crosstex GPRN95* | VHP 7% | 2 | 99.62 | 99.03 | No control provided | Not assessed | Report 9 |
Crosstex GPRN95* | VHP 7% | 3 | 99.28 | 97.85 | No control provided | Not assessed | Report 9 |
Crosstex GPRN95* | VHP 7% | 4 | 99.22 | 96.78 | No control provided | Not assessed | Report 9 |
Crosstex GPRN95* | VHP 7% | 5 | 99.13 | 96.84 | No control provided | Not assessed | Report 9 |
Gerson 1730 | VPHP | 5 | 99.65 | 99.26 | No visual degradation. <5% change in force |
≥100 | Report 17 |
Gerson 1730 | VPHP | 10 | 99.44 | 97.77 | No visual degradation >15% change in force |
≥100 | Report 17 |
Gerson 1730 | VPHP | 20 | 99.48 | 99.20 | No visual degradation >5% change in force |
≥100 | Report 17 |
Gerson 1740 | VPHP | 5 | 99.35 | 98.82 | No visual degradation >10% change in force |
Min: 88 Max: ≥100 |
Report 17 |
Gerson 1740 | VPHP | 10 | 99.40 | 98.47 | No visual degradation >5% change in force |
Min: 68 Max: ≥100 |
Report 17 |
Gerson 1740 | VPHP | 20 | 99.36 | 98.58 | No visual degradation >5% change in force |
≥100 | Report 17 |
Halyard 46727* | VHP | 5 | 99.33 | 97.89 | No visual degradation >15% change in force |
≥100 | Report 21 |
Halyard 46727 | Bioquell HPV, >480 ppm HPV at 900 sec/cycle | 5 | 99.74 | 98.44 | No visual degradation <5% change in force |
≥100 | Report 43 |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 10 | 99.46 | 96.15 | >5% change in force Discolored nosepiece foam. |
≥100 | Report 34 |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 20 | 99.55 | 85.60 | <5% change in force Discolored nosepiece foam |
Min: 44 Max: ≥100 |
Report 34 |
Makrite 9500-N95S* | H₂O₂ (35% concentration) + catalyst | 30 | 99.54 | 98.10 | >10% change in force Discolored nosepiece foam |
Min: 28 Max: 88 |
Report 34 |
Moldex 1512* | VHP 7% | 1 | 98.79 | 97.91 | No control provided | ≥100 | Report 9 |
Moldex 1512* | VHP 7% | 2 | 99.05 | 98.74 | No control provided | ≥100 | Report 9 |
Moldex 1512* | VHP 7% | 3 | 98.59 | 97.80 | No control provided | ≥100 | Report 9 |
Moldex 1512* | VHP 7% | 4 | 98.78 | 97.76 | No control provided | ≥100 | Report 9 |
Moldex 1512* | VHP 7% | 5 | 98.74 | 97.55 | No control provided | ≥100 | Report 9 |
Moldex 1512 | VPHP | 20 | 99.53 | 97.56 | No visual degradation >15% change in force |
≥100 | Report 14 |
Moldex 1512 Medium* | VPHP | 5 | 99.12 | 96.25 | No visual degradation > 10% change in force |
≥100 | Report 2 |
Moldex 1512 Medium | VPHP | 10 | 99.40 | 95.19 | No visual degradation ≤ 10% change in force |
≥100 | Report 7 |
Moldex 2200 | VPHP | 20 | 99.46 | 99.38 | No visual degradation <5% change in force |
Min: 90 Max: ≥100 |
Report 14 |
Moldex 2200 M/L* | VPHP | 5 | 99.25 | 97.54 | No visual degradation ≤ 5% change in force |
Min: 34 Max: ≥100 |
Report 2 |
Moldex 2200 M/L | VPHP | 10 | 98.76 | 94.27 | No visual degradation ≤ 10% change in force |
≥100 | Report 7 |
Sperian N1105 | VPHP | 10 | 99.88 | 99.74 | No visual degradation >10% change in force |
≥100 | Report 13 |
Sperian N1105 | VPHP | 20 | 99.94 | 99.75 | No visual degradation >10% change in force |
Min: 58 Max: ≥100 |
Report 14 |
Sperian N1105 M/L | VPHP | 5 | 99.85 | 99.54 | No visual degradation ≤ 20% change in force |
≥100 | Report 7 |
Sperian N1125 | VPHP | 10 | 99.83 | 99.24 | No visual degradation >5% change in force |
Min: 89 Max: ≥100 |
Report 14 |
Sperian N1125 | VPHP | 20 | 99.85 | 99.33 | No visual degradation >5% change in force |
Min: 58 Max: ≥100 |
Report 14 |
Sperian N1125 M/L | VPHP | 5 | 99.88 | 99.13 | No visual degradation ≤ 15% change in force |
Min: 99 Max: ≥100 |
Report 7 |
Sperian One-Fit | VPHP | 5 | 99.75 | 98.90 | No visual degradation ≤ 5% change in force |
Not assessed | Report 7 |
Sperian One-fit | VPHP | 10 | 99.80 | 97.01 | No visual degradation >15% change in force |
Not assessed | Report 14 |
Sperian One-fit | VPHP | 20 | 99.85 | 99.42 | No visual degradation >15% change in force |
Not assessed | Report 14 |
* Testing for this model did not meet the requirements of the Decontaminated Respirator Assessment Plan because an insufficient number of samples were provided to NIOSH for 1 or more tests. Therefore, results are posted for the number of samples available. Please see results report for actual number of samples evaluated for filtration efficiency, manikin fit, and strap performance.