At a glance
Table 10 from the Guideline for Disinfection and Sterilization in Healthcare Facilities (2008).
Table
Factors
Effect
CleaningA
Failure to adequately clean instrument results in higher bioburden, protein load, and salt concentration. These will decrease sterilization efficacy.
BioburdenA
The natural bioburden of used surgical devices is 100 to 103 organisms (primarily vegetative bacteria), which is substantially below the 105-106 spores used with biological indicators.
Pathogen type
Spore-forming organisms are most resistant to sterilization and are the test organisms required for FDA clearance. However, the contaminating microflora on used surgical instruments consists mainly of vegetative bacteria.
ProteinA
Residual protein decreases efficacy of sterilization. However, cleaning appears to rapidly remove protein load.
SaltA
Residual salt decreases efficacy of sterilization more than does protein load. However, cleaning appears to rapidly remove salt load.
Biofilm accumulationA
Biofilm accumulation reduces efficacy of sterilization by impairing exposure of the sterilant to the microbial cell.
Lumen length
Increasing lumen length impairs sterilant penetration. May require forced flow through lumen to achieve sterilization.
Lumen diameter
Decreasing lumen diameter impairs sterilant penetration. May require forced flow through lumen to achieve sterilization.
Restricted flow
Sterilant must come into contact with microorganisms. Device designs that prevent or inhibit this contact (e.g., sharp bends, blind lumens) will decrease sterilization efficacy.
Device design and construction
Materials used in construction may affect compatibility with different sterilization processes and affect sterilization efficacy. Design issues (e.g., screws, hinges) will also affect sterilization efficacy.
Modified from Alfa and Rutala. 470, 825
- Updates
- Authors
- Summary of Recommendations
- Introduction
- Rational Approach
- Healthcare Equipment
- Efficacy Factors
- Cleaning
- Disinfection
- Chemical Disinfectants
- Miscellaneous Inactivating Agents
- Regulatory Framework
- Neutralization of Germicides
- Sterilization
- Steam Sterilization
- Flash Sterilization
- Low-Temperature Sterilization
- Ethylene Oxide "Gas" Sterilization
- Hydrogen Peroxide Gas Plasma
- Peracetic Acid Sterilization
- Microbicidal Activity of Low-Temperature Sterilization
- Bioburden of Surgical Devices
- Cleaning Effect on Sterilization
- Other Sterilization Methods
- Sterilizing Practices
- Reuse of Single-Use Medical Devices
- Conclusion
- Performance Indicators
- Glossary
- References
- Executive Summary
- Tables and Figure
- Sterilization and Disinfection Methods
- Properties of an ideal disinfectant
- Epidemiologic Evidence
- Resistance of Microorganisms
- Comparison of Selected Chemicals
- Chemical Advantages and Disadvantages
- Sterilization Advantages & Disadvantages
- Minimum Steam Sterilization Cycle Times
- Flash Steam Sterilization Parameters
- Ideal Low-Temperature Sterilization
- Sterilization Efficacy
- Evaluation of Microbicidal Activity
- Positive Biological Indicator
- Factor only relevant for reused surgical/medical devices