Table 8 from the Guideline for Disinfection and Sterilization in Healthcare Facilities (2008).
Table
Type of Sterilizer | Load configuration | Temperature | Time |
---|---|---|---|
Gravity displacement | Nonporous items only (i.e., routine metal instruments, no lumens) | 132°C (270°F) | 3 minutes |
Gravity displacement | Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together | 132°C (270°F) | 10 minutes |
Prevacuum Dynamic-air-removal | Nonporous items only (i.e., routine metal instruments, no lumens) | 132°C (270°F) | 3 minutes |
Prevacuum, Dynamic-air-removal | Nonporous and porous items (e.g., rubber or plastic items, items with lumens) sterilized together | 132°C (270°F) | 4 minutes |
Steam-flush pressure-pulse | Nonporous or mixed nonporous/porous items | 132°C (270°F) Manufacturers’ instruction |
4 minutes |
Modified from Association for the Advancement of Medical Instrumentation. 812, 819