Manufacturer's Recall of Rapid Cartridge Assay Kits on the Basis of False-Positive Cryptosporidium Antigen Tests --- Colorado, 2004

The Colorado Department of Public Health and Environment (CDPHE) has determined that a fourfold increase in the number of reported cryptosporidiosis cases in Colorado during January--February 2004 might be attributed primarily to false-positive test results. Since January 1, 2004, a total of 13 in-state cases and one out-of-state case were reported to CDPHE. During the previous 7 years, an average of three cases were reported during January--February. In eight of 14 patients, rapid testing was performed by using the ImmunoCard STAT!® Cryptosporidium/Giardia Rapid Assay (Meridian Bioscience, Inc., Cincinnati, Ohio). This assay is a solid-phase qualitative immunochromatographic assay designated to detect and distinguish between Giardia intestinalis (lamblia) and Cryptosporidium parvum in aqueous extracts of human fecal specimens. Seven of these samples were tested by using lot no. 081093 (expires August 11, 2004). Of the seven samples that tested positive initially for Cryptosporidium with this lot number, four were retested by using other, more specific tests. One patient sample was positive by direct microscopy, one was negative by direct microscopy, and two were negative by direct fluorescent-antibody testing, suggesting that results for three of the four samples were false positive. The results of testing for Giardia intestinalis (lamblia) with these kits are unclear. Several other states have noted increases in the number of reported cryptosporidiosis cases that also might be associated with use of these rapid assays.

Meridian Bioscience, Inc., has voluntarily recalled two lots (lot no. 081077 [expires July 11, 2004] and lot no. 081093) from laboratories. CDC recommends reconfirmation of positive test results obtained with ImmunoCard STAT!® rapid assays from these lots.

Reported by: A Cronquist, MPH, Colorado Dept of Public Health and Environment. MJ Beach, PhD, SP Johnston, MS, A da Silva, PhD, Div of Parasitic Diseases, National Center for Infectious Diseases, CDC. All MMWR references are available on the Internet at http://www.cdc.gov/mmwr. Use the search function to find specific articles.

-- -- -- -- -- --

Use of trade names and commercial sources is for identification only and does not imply endorsement by the U.S. Department of Health and Human Services.

-- -- -- -- -- --

References to non-CDC sites on the Internet are provided as a service to MMWR readers and do not constitute or imply endorsement of these organizations or their programs by CDC or the U.S. Department of Health and Human Services. CDC is not responsible for the content of these sites. URL addresses listed in MMWR were current as of the date of publication.

Use of trade names and commercial sources is for identification only and does not imply endorsement by the U.S. Department of Health and Human Services.


References to non-CDC sites on the Internet are provided as a service to MMWR readers and do not constitute or imply endorsement of these organizations or their programs by CDC or the U.S. Department of Health and Human Services. CDC is not responsible for the content of pages found at these sites. URL addresses listed in MMWR were current as of the date of publication.

Disclaimer   All MMWR HTML versions of articles are electronic conversions from ASCII text into HTML. This conversion may have resulted in character translation or format errors in the HTML version. Users should not rely on this HTML document, but are referred to the electronic PDF version and/or the original MMWR paper copy for the official text, figures, and tables. An original paper copy of this issue can be obtained from the Superintendent of Documents, U.S. Government Printing Office (GPO), Washington, DC 20402-9371; telephone: (202) 512-1800. Contact GPO for current prices.

**Questions or messages regarding errors in formatting should be addressed to mmwrq@cdc.gov.

Page converted: 3/4/2004

HOME  |  ABOUT MMWR  |  MMWR SEARCH  |  DOWNLOADS  |  RSSCONTACT
POLICY  |  DISCLAIMER  |  ACCESSIBILITY

Safer, Healthier People

Morbidity and Mortality Weekly Report
Centers for Disease Control and Prevention
1600 Clifton Rd, MailStop E-90, Atlanta, GA 30333, U.S.A

USA.GovDHHS

Department of Health
and Human Services

This page last reviewed 3/4/2004